Spectroscopy in the Laboratory

Astronomical Laboratory 29:137, Fall 2013
by
Philip Kaaret and Steven Spangler

Reading


Introduction

Spectroscopy is crucial to astronomy.  It is the principal diagnostic for determining the temperature and chemical composition of stars, nebulae, and galaxies.  Spectroscopy is of equal importance to physics.  No experimental evidence was of more importance to the emergence of modern physics than the existence of spectral lines of different atoms and molecules.  Today, spectroscopy continues to be one of the most important experimental or observational measurement in both astronomy and physics. 

In this class, we will have the pleasure of measuring and analysing spectra. We will use the same ST-402XME camera used in the previous exercises as the detector of our spectrograph.  The equipment which actually makes up the spectrograph (slit, internal optics, and most importantly, diffraction grating) is provided by a unit which connects to the front of the ST-402XME. The instrument is called the Deep Space Spectrograph DSS-7.  This unit is also manufactured by the Santa Barbara Instrumentation Group (SBIG).  The SBIG company designed this spectrograph to be mounted on a telescope. 

Today, we will set up the DSS-7 with the ST-402 in the laboratory and get it working.  We will then calibrate it by taking spectra of gas discharges of hydrogen, helium, etc.

This exercise will be much more meaningful if you know what is going on.  Be sure to read sections 6.1-6.5 in Handbook of CCD Astronomy before coming to class.  Also, the manual for the DSS-7 spectrograph has much worthwhile tutorial material as well as indispensible material on the construction and operation of the DSS-7. You should particularly look at Figures 1, 5, and 7 of the DSS-7 manual. An important characteristic of the DSS-7 is its wavelength range.  As stated in the manual, it is nominally 400-800 nanometers (or 4000-8000 Angstroms, in a unit more traditionally used in astronomy).  

Equipment

System Setup


Centering and Focusing on the Slit

The goal is to make sure that the internal optics of the spectrograph are focused on the slit and that the slit images are falling in the middle of the CCD chip (instead of off to the side).

          


Let's take a moment to discuss the spectrum images.  The image of the spectrum of a hydrogen lamp is shown below.  The horizontal axis in the image below is the 'dispersion axis', i.e. the axis along which the photons are dispersed by the grating according to their wavelength.  Thus, position along the horizontal axis is a measure of wavelength (which we will calibrate in detail below).

The vertical axis is un-dispersed.  The grating essentially acts like a mirror for reflection in this axis.  Thus, the vertical axis is actually a spatial axis.  A vertical slice across the spectral image through a bright spectral line will give a one-dimensional image of the source along an axis aligned with the slit.  For the hydrogen lamp, the light source fills the whole slit, so one sees the whole outline of the slit in the bright lines.  In contrast, a star will fill only a few pixels of the vertical extent of the slit.

         
Sometimes spectra images are binned along the vertical direction.  Typically one integrates a stellar spectrum across the whole width of the stellar image, usually several pixels.  When acquiring spectra, CCDOps has an option to do this binning for you.  Setting 'Vertical Binning' to a value greater than one, simply combines that number of adjacent vertical pixels.  To set the vertical binning for dark frames, do Camera/Setup and look at the 'CCD Setup' part of the dialog box.  It should be fine to set 'Vertical Binning' to a small whole number to combine vertical sets of pixels and allow shorter exposures times.  However, it seems that CCDOps ignores 'Vertical Binning' when taking dark frames, so keep 'Vertical Binning' set to 1.  It is important to have a set of dark frames that can be directly subtracted from the spectrum images; the dark frames should have the same vertical binning and exposure time.

We are now ready to take data.
 

Data Taking and Measurements

The main goal of this lab exercise is to calibrate the spectrograph.  You will now take spectra of hydrogen, helium, and neon lamps and blue and red LEDs for calibration purposes.  We will need to have the lights off in the lab for this part.

Gratings

When light is normally incident on a grating, the diffracted light will have maxima at angles θm given by:

 d \ \sin{\theta_m} = m \lambda

where d is the spacing of the rulings on the grating, m is the order number and can be any integer, λ is the wavelength of the light.  This is the so-called 'grating equation'.  From inspection of Figure 1 in the DSS-7 manual (or inspection of the DSS-7 itself with the panel removed), one can see that the DSS-7 is designed so that the light is normally incident on the grating (when used in spectroscopic mode).  The DSS-7 grating operates in first order when used in spectroscopic mode.  Thus m = 1.


Calibrating the Spectrograph

In principle, one could measure the d and then trace the path of rays through the DSS-7 and ST-402 to find the angle θ corresponding to each pixel in an image.  In practice, this is too difficult and fraught with error, so one, instead, observes a light source, usually a gas discharge lamp, that produces several emission lines of known wavelength and uses those to determine the relationship between pixel number and wavelength of spectral lines for the spectrograph.  This is called 'calibrating the spectrograph'.

To do this, one must first identify the various lines in each spectrum.

You will now calibrate your spectrograph using Python and your hydrogen spectrum with its newly identified lines. 

In general, there can be a complex relation between pixel number p and diffraction angle θ that depends on the optical properties of the spectrograph.  If the diffraction angles are small, θ << 1, then the relation between pixel number p and diffraction angle θ can be approximated as p = aθ + b.  We will use a linear relationship.  Note that since there are only two strong lines in the hydrogen spectrum, we don't actually have enough information to derive a more complex calibration curve.  In astronomical research, much larger line sets, often 20 to 40 lines or more, are used to correct for non-linearities in the pixel number versus wavelength relation.  The software that comes with the DSS-7 also only uses two lines and the manual states that the calibration is good to 0.1 pixels.

Having now calibrated our spectrograph, we can test the accuracy of the calibration. 

For extra credit, you can repeat this exercise for your neon spectrum.  A list line for neon is available here: http://physics.nist.gov/PhysRefData/Handbook/Tables/neontable2.htm (not just any line list, the official US line list).  If you are planning on using the DSS-7 for your research project, you should think about how you can use the whole set of line spectra to improve the calibration.